Pesola Replacement Inner Tube of MFT-Scale 43000/5
This replacement part only fits the MFT-scale 43000/5 for myofunctional therapy (logopedia). Despite of shock absorbing elements, the strain of recoil after release at max load of several 100 cycles may cause a fatigue failure of the interior tube.
Instruction for installation of replacement part
Guide
- Turn the setscrew counterclockwise until the spring with the interior tube is no longer held in the scale.
 - Pull out defective inner tube (possibly press the elastic marker into the slit).
 - Insert the new replacement part 41.43005 tube ahead into the outer aluminum tube.
 - Therefore press in the marker slightly.
 - Insert the plastic inner tube further more and turn it simultaneously until the marker clicks into the slit.
 - Hold the scale upside down and turn the setscrew clockwise until the zero point is readjusted (scale now in measuring position)
 
How to avoid damage
- Only use one drag pointer below the marker. Otherwise the recoil might damage the marker. The second drag pointer is for replacement purposes.
 - Do not use any solvents for disinfecting purposes. The plastic elements might turn brittle.
 
Maximum conditions for sterilisation
- In autoclave at 120°C at 1 bar for 45 minutes
 - or 15 h in acetylene
 
Attention: synthetic parts are incompatible with high concentrations of chlorine and its compounds.